Z-0173-2022Class II
C315-HIS Delivery Catheter
Medtronic, Inc. / initiated 2021-09-09 / Open, Classified
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of C315-HIS Delivery Catheter on . FDA classified it as Class II and gives the reason as "Medtronic is recalling three lots of unused C315HIS Delivery Catheters due to reports of an inability to pass a lead through the catheter". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0173-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- C315-HIS Delivery Catheter
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Medtronic is recalling three lots of unused C315HIS Delivery Catheters due to reports of an inability to pass a lead through the catheter.
- FDA root cause
- Under Investigation by firm
- Action
- Written notification letter titled, "MEDICAL DEVICE RECALL C315HIS DELIVERY CATHETER" was sent to all consignees on 09/09/2021 via 2-day UPS delivery. The letter instructs the consignee to take the following actions: "Check your inventory for the catheters identified in Table 1 above. If the product has been consumed, please complete the enclosed Customer Confirmation Form and email to [email protected]. No further actions are required. " If there is unused identified product in your inventory, quarantine and return it to Medtronic. Contact Medtronic Customer Service at xxxxxxx to initiate a product return. Your local Medtronic Representative can assist you in the return of this product. o Note: Root cause is under investigation for this issue. Return of the identified product will aid the investigation. Please complete the enclosed Customer Confirmation Form and email to [email protected]." The letter also instructs the consignee to share the notice with all those who need to be made aware with their organization or to any organization where the potentially affected products have been transferred.
- Quantity in commerce
- 591 units
- Distribution
- Distributed internationally to Australia, Brunei Darussalam, China, Germany, Hong Kong, Indonesia, Malaysia, Netherlands, Singapore, and Taiwan.
- Product codes and lots
- GTIN: 00763000160180; Model: C315HIS02 (Serial Number/Lot Number: 0010567341), C315HIS03 (Serial Number/Lot Number: 10568899), and C315HIS04 (Serial Number/Lot Number: 10568900)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0173-2022 | Class II | Ongoing | Voluntary: Firm initiated |