Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0074-2022Class II

ATTUNE Revision Tibial Augment Size 7/8 5mm- intended for knee joint replacement Part Code:152307001

Depuy Orthopaedics Inc. / initiated 2021-09-09 / Terminated / root cause: packaging

DePuy Orthopaedics, Inc. began a recall of ATTUNE Revision Tibial Augment Size 7/8 5mm- intended for knee joint replacement Part Code:152307001 on . FDA classified it as Class II and gives the reason as "Packaging may include screws that are not intended to be used with the 5mm augments, may cause delay in surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0074-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
ATTUNE Revision Tibial Augment Size 7/8 5mm- intended for knee joint replacement Part Code:152307001
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Packaging may include screws that are not intended to be used with the 5mm augments, may cause delay in surgery
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
DePuy (Ireland) issued Urgent Medical Device Recall (Removal)letter via email to US Sales Offices/Distributors and US Hospitals on 9/9/21. Recall letters will be emailed to US Sales Offices/Distributors. US Hospitals will be contacted by sales consultants. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have the subject lot and quarantine the product. 2. Return the subject product(s) using the normal returns process. Work with your sales consultant to return subject product(s) and to replace. To receive replacement product(s) or reimbursement, customers must return the products subject to this removal. 3. Review, complete all fields, sign, and return the attached business response form (BRF) on page 2 of this letter to [email protected] within three (3) business days of receipt of this Removal Letter. Please include in the email subject: FA 2010228 - ATTUNE Revision Tibial Augment Incorrect Screw Size Packaged. 4. Forward this Removal Letter to any personnel in your facility who need to be informed. 5. If any of the product(s) subject to this Removal have been forwarded to another facility, contact that facility and provide a copy of this Removal Letter to the relevant personnel. 6. Post a copy of this Removal Letter in a visible area for awareness of this Removal. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.
Quantity in commerce
10 units
Distribution
US Nationwide distribution in the states of CA, LA, MN, MO, NC, NE, OR, PA, WA, WI.
Product codes and lots
Lot Number: J6304Z GTIN: 10603295057642

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0074-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0074-2022Class IITerminatedVoluntary: Firm initiated