Z-0174-2022Class II

Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic...

Xstrahl Limited / initiated 2021-04-26 / Terminated / root cause: software

Xstrahl Limited began a recall of Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy System; (2)... on . FDA classified it as Class II and gives the reason as "If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0174-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Xstrahl Limited
Product
Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy System; (2) Xstrahl 150 Electronic Brachytherapy - 150kV Superficial X-Ray Therapy System; (3) Xstrahl 200 - 225kV Superficial / Orthovoltage X-Ray Therapy System; (4) Xstrahl 300 - 300kV Orthovoltage X-Ray Therapy System; (5) Xstrahl X 80 Photoelectric Therapy System.
Product code
JAD System, Therapeutic, X-Ray
Reason for recall
If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued an Urgent Field Safety Notice dated 4/26/2021 via email on 4/26/2021. The notice explained the issue, impact of the issue, and provided actions to be taken for Concerto 2.x 2 Beam Plans Only, although all customers with the affected software were issued the safety notice. The consignee was informed that until an update to their affected software version is installed, they should follow the actions listed in the letter.
Quantity in commerce
25 systems in the U.S.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX. The countries of Austria, Germany, and Switzerland.
Product codes and lots
V2.0, V2.1, and V2.2 when 2 opposing beam treatment plans are used.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0174-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0174-2022Class IITerminatedVoluntary: Firm initiated