Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0209-2022Class II

HENRY SCHEIN Safety Blood Collection Needle 22Gx 1 1/4 Inch, REF 570-2068, 100pcs/box 1000pcs/ctn...

Zhejiang Kindly Medical Devices Co., Ltd. / initiated 2021-08-24 / Terminated / root cause: manufacturing process

Zhejiang Kindly Medical Devices Co., Ltd. began a recall of HENRY SCHEIN Safety Blood Collection Needle 22Gx 1 1/4 Inch, REF 570-2068, 100pcs/box 1000pcs/ctn 350*210*350mm on . FDA classified it as Class II and gives the reason as "Potential for difficulty in safety mechanism activation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0209-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zhejiang Kindly Medical Devices Co., Ltd.
Product
HENRY SCHEIN Safety Blood Collection Needle 22Gx 1 1/4 Inch, REF 570-2068, 100pcs/box 1000pcs/ctn 350*210*350mm
Product code
FMI Needle, Hypodermic, Single Lumen
Reason for recall
Potential for difficulty in safety mechanism activation.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 8/24/2021, Zhejiang Kindly Medical Devices Co., Ltd. issued Urgent Medical Device Safety Alert notices to two customers. On 9/17/2021, issued a second notice to customers dated 8/24/2021. Action to be Taken by User: 1 - Advise the user to lower the angle between safety protective cap and needle tube by slight pressing in the direction of activation, then using a thumb or hard surface to activate the safety protective cap and lock the needle tube in the IFU, then together with corresponding illustration. 2 - Request the user to strictly follow the guidance of the IFU.
Distribution
U.S. Nationwide distribution.
Product codes and lots
Lot Numbers / Expiration Date: 20201030 / Exp. Date 2025.10.29, 20201210 / Exp. Date 2025.12.09.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0209-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0204-2022Class IITerminatedVoluntary: Firm initiated
Z-0205-2022Class IITerminatedVoluntary: Firm initiated
Z-0206-2022Class IITerminatedVoluntary: Firm initiated
Z-0207-2022Class IITerminatedVoluntary: Firm initiated
Z-0208-2022Class IITerminatedVoluntary: Firm initiated
Z-0209-2022Class IITerminatedVoluntary: Firm initiated