Z-0245-2022Class II

Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434

Medicalplastic / initiated 2021-06-17 / Open, Classified / root cause: manufacturing process

Medicalplastic S.R.L. began a recall of Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434 on . FDA classified it as Class II and gives the reason as "Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0245-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medicalplastic
Product
Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm notified its sole consignee of the issue and its intent to recall by email on 06/17/2021. The recalled washers will be replaced.
Quantity in commerce
581 units
Distribution
Memphis, TN
Product codes and lots
Lot # 20LP02813, Serial #D147 to D149; Lot # 20LP02812, Serial # D115 to D119 ; Lot # 20LP02816, Serial # D099 to D114 ; Lot # 20HP01926, Serial # D059 to D065 ; Lot # 20LP02819, Serial # D175 to D198 ; Lot # 20LP02818, Serial # D129 to D146 ; Lot # 20LP02815, Serial # D120 to D128 ; Lot # 20MP03078, Serial # D374 ; Lot # 20HP01927, Serial # D278 to D282 ; Lot # 20MP03037, Serial # D294 to D296 ; Lot # 20LP02827, Serial # D199 to D258 ; Lot # 20 MP03036, Serial # D288 to D293 ; Lot # 20MP03077, Serial # D370 to D373 ; Lot # 20MP03035, Serial # D259 to D277 ; Lot # 20MP03041, Serial # D283 to D287 ; Lot # 20MP03049, Serial # D343 to D369 ; Lot # 20MP03048, Serial # D297 to D342 ; Lot # 20MP03123, Serial # D375 to D459 ; Lot # 20HP01928, Serial # D460 to D474 ; Lot # 20MP03126, Serial # D482 to D491 ; Lot # 20MP03125, Serial # D492 to D522 ; Lot # 20MP03124, Serial # D475 to D481 ; Lot # 21BP00265, Serial # D523 to D574 ; Lot # 21BP00266, Serial # D626 to D672 ; Lot # 21BP00271, Serial # D588 to D625 ; Lot # 21BP00272, Serial # D575 to D587 ;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0245-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0245-2022Class IIOngoingVoluntary: Firm initiated