Z-0125-2022Class II

Atellica IM 1600 Analyzer, SMN 11066000

Siemens Healthcare Diagnostics Inc. / initiated 2021-09-28 / Open, Classified

Siemens Healthcare Diagnostics, Inc began a recall of Atellica IM 1600 Analyzer, SMN 11066000 on . FDA classified it as Class II and gives the reason as "Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result, customers...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0125-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Atellica IM 1600 Analyzer, SMN 11066000
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result, customers at these altitudes may observe an increased number of Reagent Volume Check Errors.
FDA root cause
Under Investigation by firm
Action
On September 27, 2021, the firm initiated notifications to affected customers. Customer distribution included all customers located in a city at an altitude of 350 meters (1148 feet) or higher. Customers were informed that Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350 (m) or 1148 (ft) above sea level. The letters provide directions for mitigating the potential for leakage. If you have any questions, contact your local Siemens Healthineers technical support representative.
Quantity in commerce
1733
Distribution
Domestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX UT VA WA WI WY. Foreign distribution worldwide.
Product codes and lots
All units.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0125-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0124-2022Class IIOngoingVoluntary: Firm initiated
Z-0125-2022Class IIOngoingVoluntary: Firm initiated