Z-0183-2022Class I
Ellume COVID-19 Home Test
Ellume Ltd / initiated 2021-10-01 / Open, Classified / root cause: manufacturing process
ELLUME LTD began a recall of Ellume COVID-19 Home Test on . FDA classified it as Class I and gives the reason as "Due to a higher rate of false positive test results". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0183-2022
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ellume Ltd
- Product
- Ellume COVID-19 Home Test
- Product code
- QKP Coronavirus Antigen Detection Test System.
- Reason for recall
- Due to a higher rate of false positive test results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 10/01/2021, the Recalling Firm communicated an "URGENT: MEDICAL DEVICE RECALL" Notification letter via email to its: -Distributors/retailer -Customers/users that include customers that have used the products within 2 weeks and beyond 2 weeks. The communications informed customer that the Recalling Firm has received a number of complaints stating false positive test results for it COVID-19 Home Test kits. If consumers have used an affected product before 09/17/2021 and tested positive are advised that their test results could have been incorrect. For Distributor and Retailers, they are instructed to: 1. Remove affected products from their shelves and cease sales and distibution. 2. Quarantine the affected products immediately 3. Contact the Recalling Firm's sales representatives for further instructions on the disposal of affected product. 4.Complete and return the attached Acknowledgement Form acknowledging their receipt of the Recall Notification and confirming that they have returned or destroyed the affected product. Return Acknowledgement Form to the Recalling Firm at [email protected] 5. If the affected products have been further distributed, notify any accounts or additional locations and share the Recall Notification with them 6. Immediately notify the Recalling Firm of any accounts or additional locations that may have received the affected product. For Consumers that have used the affected products: Users that have used the affected products within the last two weeks are being informed and instructed to: 1. If they have received a positive test result, this result may be incorrect and they should contact a healthcare professional immediately and notify them of this recall and ask to obtain a confirmatory test (i.e., molecular or RT-PCR). 2. If they have already had a confirmatory test that confirms they DO have COVID-19, disregard this letter and follow their healthcare professional's advice and CDC guidance on self-isolation,
- Quantity in commerce
- 427,994 kits
- Distribution
- U.S.: MN, NY, PA, RI, UT, WA. O.U.S.: Not provided
- Product codes and lots
- Catalogue number I-SRS-C-01 Master Lot Numbers: 21047-4; 21047-5; 21089-1; 21117-1; PF06Z-H; 21099-1; 21124-1; 21125-1; PF03X-H; PF057-H; PF05W-H; PF069-H; PF06E-H; PF06N-H; PF06Z-H; PG080-H; PG08H-H; PH08X-H;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0183-2022 | Class I | Ongoing | Voluntary: Firm initiated |