Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned...
Windstone Medical Packaging, Inc. began a recall of Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned Medical Solutions Custom Convenience Kits... on . FDA classified it as Class I and gives the reason as "Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the syringe". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0119-2022
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Windstone Medical Packaging, Inc.
- Product
- Monoject 0.9% Sodium Chloride, Flush Prefilled Syringe, 10 mL, REF: 8881570121 inside the following Aligned Medical Solutions Custom Convenience Kits: AMS-9041CP Leaderflex Insertion Kit with Ultrasound, AMS-9046CP-1 Insertion Tray-RX, AMS8939A Universal Procedure Pack w/Split Drape, AMS9957A Port Insertion Pack, and AMS12149 Procedure Pack
- Product code
- NGT Saline, Vascular Access Flush
- Reason for recall
- Monoject 0.9% Sodium Chloride Flush Syringe Manufactured by Cardinal Health and placed into kits. Potential for the plunger to draw back after air has been expelled and reintroduce air back into the syringe.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 08/19/2021, recall notices were emailed to customers. Customers were asked to do the following: Identify remaining affected kits and follow the over-labeling instructions. The sticker should be placed in such a way that it would be immediately noticed but should not cover up or conceal any pertinent information such as Pack Number, Pack Name, Lot Number or Expiration Date. Sticker states in part: A recall of the prefilled syringe contained in this kit was announced. All other kit components are not affected. 1. At the time the kit is opened for use any prefilled syringe manufactured by Cardinal Health should be identified and set aside 2. The recalled syringe should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. Complete and return the reply form. If product was further distributed, notify your customers of the product recall Customers with questions or concerns can contact the recalling firm at 407-638-9924, www.alignedmedicalsolutions.com There is a recall notice on the website: http://www.alignedmedicalsolutions.com/ams_recall_08192021.htm
- Quantity in commerce
- 9,378
- Distribution
- FL and PA
- Product codes and lots
- Syringe Lots: All Lots manufactured and distributed between July 1, 2019 and July 2021. Kit Model/Lots: AMS-9041CP/ 136833; AMS-9046CP-1/ 137794; AMS8939A/ 140390, 141088, 141096, 141100, 141102, 146039, 146349, 147902, 147903, 148414, 148452, 148640, 149132, 149135, 149485, 141087, 141097, 145615, 146351, 146766, 146768, 146905, 146908, 148638, 149137, 149138, 149484, 149486, 141101, 148410, 148411, 148639, 149133, 141098, 146037, 148409, 148413, 148450, 149131, 149134, 149136, 148412, 148451, 148636, 141717 AMS9957A/ 147773; AMS12149/ 145718, 146897, 148612