Z-0223-2022Class II

Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer...

Cook Incorporated / initiated 2021-10-08 / Terminated / root cause: manufacturing process

Cook Inc. began a recall of Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip... on . FDA classified it as Class II and gives the reason as "Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0223-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle
Product code
N/A
Reason for recall
Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On October 8, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were instructed to take the following actions: 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of these products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected]. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Quantity in commerce
4789 (OUS only)
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Product codes and lots
All Lots  UDI (01)00827002192614(17)211010(10)9226180 (01)00827002192614(17)211015(10)9234901 (01)00827002192614(17)211015(10)9235946 (01)00827002192614(17)211015(10)9235947 (01)00827002192614(17)211017(10)9242172 (01)00827002192614(17)211120(10)9326411 (01)00827002192614(17)211120(10)9326412 (01)00827002192614(17)211121(10)9328622 (01)00827002192614(17)211130(10)9347132 (01)00827002192614(17)211203(10)9351027 (01)00827002192614(17)211204(10)9354000 (01)00827002192614(17)211210(10)9370544 (01)00827002192614(17)211220(10)9396895 (01)00827002192614(17)220104(10)9419991 (01)00827002192614(17)220109(10)9431068 (01)00827002192614(17)220109(10)9431069 (01)00827002192614(17)220124(10)9472321 (01)00827002192614(17)220128(10)9480776 (01)00827002192614(17)220201(10)9492814 (01)00827002192614(17)220201(10)9492815 (01)00827002192614(17)220201(10)9492816 (01)00827002192614(17)220214(10)9524330 (01)00827002192614(17)220222(10)9545708 (01)00827002192614(17)220222(10)9545720 (01)00827002192614(17)220304(10)9568083 (01)00827002192614(17)220304(10)9568088 (01)00827002192614(17)220311(10)9583267 (01)00827002192614(17)220319(10)9604183 (01)00827002192614(17)220319(10)9604184 (01)00827002192614(17)220322(10)9611406 (01)00827002192614(17)220322(10)9613905 (01)00827002192614(17)220325(10)9617635 (01)00827002192614(17)220328(10)9626175 (01)00827002192614(17)220328(10)9626176 (01)00827002192614(17)220328(10)9626177 (01)00827002192614(17)220402(10)9638768 (01)00827002192614(17)220404(10)9643348 (01)00827002192614(17)220404(10)9643349 (01)00827002192614(17)220409(10)9654385 (01)00827002192614(17)220409(10)9654513 (01)00827002192614(17)220409(10)9654514 (01)00827002192614(17)220409(10)9655757 (01)00827002192614(17)220510(10)9730630 (01)00827002192614(17)220516(10)9743413 (01)00827002192614(17)220516(10)9743414 (01)00827002192614(17)220522(10)9760371 (01)00827002192614(17)220611(10)9802149 (01)00827002192614(17)220620(10)9826728 (01)00827002192614(17)220620(10)9826729 (01)00827002192614(17)220627(10)9840896 (01)00827002192614(17)220628(10)9844564 (01)00827002192614(17)220701(10)9850898 (01)00827002192614(17)220822(10)9966055 (01)00827002192614(17)220826(10)9971573 (01)00827002192614(17)220826(10)9971744 (01)00827002192614(17)220910(10)10002117 (01)00827002192614(17)220917(10)10016709 (01)00827002192614(17)220919(10)10025480 (01)00827002192614(17)220923(10)10031245 (01)00827002192614(17)221023(10)10102218 (01)00827002192614(17)221023(10)10102219 (01)00827002192614(17)221024(10)10105936 (01)00827002192614(17)221024(10)10105937 (01)00827002192614(17)221113(10)10148679 (01)00827002192614(17)221114(10)10151314 (01)00827002192614(17)221118(10)10157368 (01)00827002192614(17)221121(10)10166249 (01)00827002192614(17)221219(10)10224967 (01)00827002192614(17)221219(10)10224968 (01)00827002192614(17)230204(10)10307597 (01)00827002192614(17)230212(10)10324494 (01)00827002192614(17)230220(10)10342526 (01)00827002192614(17)230220(10)10342529 (01)00827002192614(17)230421(10)13145971 (01)00827002192614(17)230421(10)13145972 (01)00827002192614(17)230514(10)13193150 (01)00827002192614(17)230514(10)13193151 (01)00827002192614(17)230609(10)13235106 (01)00827002192614(17)230609(10)13235107 (01)00827002192614(17)230622(10)13260302 (01)00827002192614(17)230714(10)13304618 (01)00827002192614(17)230716(10)13311621 (01)00827002192614(17)230716(10)13311622 (01)00827002192614(17)230716(10)13311623 (01)00827002192614(17)231009(10)13481714 (01)00827002192614(17)231009(10)13481715 (01)00827002192614(17)231112(10)13553814 (01)00827002192614(17)231202(10)13602045 (01)00827002192614(17)231202(10)13602046 (01)00827002192614(17)231202(10)13602047 (01)00827002192614(17)231204(10)13606368 (01)00827002192614(17)231204(10)13606369 (01)00827002192614(17)231204(10)13606370 (01)00827002192614(17)231204(10)13606371 (01)00827002192614(17)240120(10)13702632 (01)00827002192614(17)240120(10)13702633 (01)00827002192614(17)240120(10)13702634 (01)00827002192614(17)240122(10)13707721 (01)00827002192614(17)240311(10)13817370 (01)00827002192614(17)240311(10)13817371 (01)00827002192614(17)240311(10)13817372 (01)00827002192614(17)220222(10)9545708X (01)00827002192614(17)191110(10)NS7433921 (01)00827002192614(17)191201(10)NS7482244 (01)00827002192614(17)191201(10)NS7482246 (01)00827002192614(17)191201(10)NS7482247 (01)00827002192614(17)200712(10)NS8059999 (01)00827002192614(17)211003(10)NS9209497 (01)00827002192614(17)211003(10)NS9209498 (01)00827002192614(17)211003(10)NS9209499 (01)00827002192614(17)211003(10)NS9209500 (01)00827002192614(17)211004(10)NS9212134

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0223-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0222-2022Class IITerminatedVoluntary: Firm initiated
Z-0223-2022Class IITerminatedVoluntary: Firm initiated