Z-0243-2022Class II
API 50 CH, Model 50300
bioMerieux, Inc. / initiated 2021-09-30 / Terminated / root cause: manufacturing process
bioMerieux, Inc. began a recall of API 50 CH, Model 50300 on . FDA classified it as Class II and gives the reason as "There is potential to provide an incorrect organism identification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0243-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- API 50 CH, Model 50300
- Product code
- JSC Culture Media, General Nutrient Broth
- Reason for recall
- There is potential to provide an incorrect organism identification.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Notification letters were issued to customers via FedEx Priority Overnight mail during the week of October 4, 2021. Customers are asked to do the following: -Discontinue use of and discard any remaining inventory of lot 1008679230. -For tests previously performed using API 50 CH Ref. 50300 lot# 1008679230, we are asking you to identify any possible incorrect identification test result due to a false positive reaction of test 34, analyze the related risks and to determine appropriate actions, if relevant. -Complete the Acknowledgement Form in Attachment A and return it to your local bioMrieux representative to confirm receipt of this notice. -Distribute this information to all appropriate personnel in the laboratory, retain a copy in your files. -Contact your local bioMrieux representative for product compensation. Consignees will perform destruction activities at their site.
- Quantity in commerce
- 1382 kits
- Distribution
- Distribution in WI, NY, and SD
- Product codes and lots
- Model 50300, Lot 1008679230 UDI: (01)03573026087401 (17)220330 (10)1008679230
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0243-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY