Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0243-2022Class II

API 50 CH, Model 50300

bioMerieux, Inc. / initiated 2021-09-30 / Terminated / root cause: manufacturing process

bioMerieux, Inc. began a recall of API 50 CH, Model 50300 on . FDA classified it as Class II and gives the reason as "There is potential to provide an incorrect organism identification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0243-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
API 50 CH, Model 50300
Product code
JSC Culture Media, General Nutrient Broth
Reason for recall
There is potential to provide an incorrect organism identification.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Notification letters were issued to customers via FedEx Priority Overnight mail during the week of October 4, 2021. Customers are asked to do the following: -Discontinue use of and discard any remaining inventory of lot 1008679230. -For tests previously performed using API 50 CH Ref. 50300 lot# 1008679230, we are asking you to identify any possible incorrect identification test result due to a false positive reaction of test 34, analyze the related risks and to determine appropriate actions, if relevant. -Complete the Acknowledgement Form in Attachment A and return it to your local bioMrieux representative to confirm receipt of this notice. -Distribute this information to all appropriate personnel in the laboratory, retain a copy in your files. -Contact your local bioMrieux representative for product compensation. Consignees will perform destruction activities at their site.
Quantity in commerce
1382 kits
Distribution
Distribution in WI, NY, and SD
Product codes and lots
Model 50300, Lot 1008679230  UDI: (01)03573026087401 (17)220330 (10)1008679230

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0243-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0243-2022Class IITerminatedVoluntary: Firm initiated