Z-0274-2022Class II

Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use...

The Binding Site Group Ltd. / initiated 2021-10-15 / Terminated / root cause: manufacturing process

The Binding Site Group, Ltd. began a recall of Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S on . FDA classified it as Class II and gives the reason as "Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0274-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Binding Site Group Ltd.
Product
Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10) REF LK018.10S
Product code
DEH Lambda, Antigen, Antiserum, Control
Reason for recall
Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 10/15/2021 (revised letter 10/25/2021), the firm sent an "Urgent Medical Device Correction" Notification to customers via email to inform them that it has identified an imprecision at top of calibration curve, where the calibration curves produced using specific lots have a higher activity than routinely expected and that there may be a decrease separation between calibrator point 5 and 6 which can be exacerbated when calibrator 6 activity reaches the optical ceiling of the analyzer and may result in an inaccurate value being reported. Customers are instructed to: Adapt the calibration curve to run as a 5-point curve with immediate effect as below 1. Stop using the 6-point calibration curve for the affected kit lots. 2. Run a 5-point calibration curve by entering values for Calibrator 1 to Calibrator 5 only as found on the Quality Control Certificate provided within the kit in ascending order on Page 1 of the Item parameters. Leave Standard 6 blank. Ensure the Auto Fill button is pressed after entering the value for Calibrator 5 on page 1. The analyser will automatically calculate and enter the revised measuring ranges on Item pages 3 and 4. 3. View Item parameter pages 3 and 4 to ensure correct value entry. 4. Before requesting calibration, view the CALIB screen and delete any existing rack position and OD information shown for STD-6 of the corresponding assay. 5. Calibration curves must only be accepted if Absorbance values increase with increasing concentration as stated in Section D.2.2 of the SPAPLUS Reference Guide (FIN012). -The expiry of the above kit lots has been reduced from 31/OCT/2022 to 30/APR/2022. Update records accordingly as the affected kit lots must not be used beyond this point. 6. Fill out and return the TSWS18 E-Back form to [email protected] or to your local The Binding Site Representative within two weeks of receiving this notification. For question/further information, contact Local Binding Site Representative or Techn
Quantity in commerce
2,254 kits
Distribution
U.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: Canada
Product codes and lots
Model Number: LK018.S UDI Code: 05051700006477 Lot Number: 472776 Model Number:LK018.10S UDI Code: 05051700012201 Lot Number: 472664

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0274-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0274-2022Class IITerminatedVoluntary: Firm initiated