Z-0290-2022Class II

Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE

Medtronic, Inc. / initiated 2021-10-15 / Open, Classified

Medtronic Vascular, Inc. began a recall of Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE on . FDA classified it as Class II and gives the reason as "During stent graft deployment, the spindle may detach from the distal end of the spindle hypotube". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0290-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
During stent graft deployment, the spindle may detach from the distal end of the spindle hypotube
FDA root cause
Under Investigation by firm
Action
On or about 10/15/2021, the firm initiated a global verbal notification to customers informing them that is the potential for a specific subset of 18Fr Stent Graft Systems to have the potential for the spindle to detach from the spindle hypotube during deployment. On 10/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-day delivery. Customers are instructed to take the following actions: 1) Return all unused affected devices to the Recalling Firm via their sales representative or calling customer service at 1-888-283-7868 or email [email protected]. 2) Complete the enclosed Customer Confirmation Form and email to [email protected]. 3) Forward this Recall notice to all those who need to be aware within their organization to include physician implanters (US only). Additionally, if any affected devices have been distributed to other organizations, forward this notice to those entities. For questions or assistance - contact their Field Representative or Customer Service at 888-283-7868 or email [email protected].
Quantity in commerce
2 devices
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of CA, FL, GA, MN, MO, NC, NJ, NM, NV, NY, PA, SC, TN, TX, VA, and WV. The countries of Austria, Belgium, Bulgaria, China, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hungary, Ireland, Italy, Japan, Jordan, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Qatar, South Africa, Spain, Sweden, Switzerland, and United Kingdom.
Product codes and lots
Model Number/Product Descripton: ENTF2525C70EE/STENT GRAFT ENTF2525C70EE ENDURANT TUBE GTIN Code: 00763000006129 Serial Numbers: V30624638 V30624639

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0290-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0290-2022Class IIOngoingVoluntary: Firm initiated
Z-0291-2022Class IIOngoingVoluntary: Firm initiated
Z-0292-2022Class IIOngoingVoluntary: Firm initiated