Artis zee ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging...
Siemens Medical Solutions USA, Inc began a recall of Artis zee ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number:10094137 on . FDA classified it as Class II and gives the reason as "If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0297-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zee ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number:10094137
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice dated 10/202/1 to customers via AX070/21/S, Siemens will update the system software via Update Instruction AX069/21/S. Letter states reason for recall, health risk and action to take: To avoid the problem, it is recommended to position the SID first and then release x-ray. Siemens will correct the software error (software VD12) via Update Instruction AX069/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. We do not consider it necessary to re-examine any patients in relation to the issues described above. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
- Quantity in commerce
- 8 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Serial Numbers: Serial 150002 150017 125700 150005 150023 150009 150015 150003 Expanded Recall 11/15/21: 147120 147119 150011
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0296-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0297-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0298-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0299-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0300-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0301-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0302-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0303-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0304-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0305-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
