Z-0300-2022Class II

Artis Q floor with software VD12- angiography systems developed for single and biplane diagnostic imaging and...

Siemens Medical Solutions USA, Inc. / initiated 2021-10-20 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Artis Q floor with software VD12- angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848280 on . FDA classified it as Class II and gives the reason as "If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0300-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis Q floor with software VD12- angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848280
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice dated 10/202/1 to customers via AX070/21/S, Siemens will update the system software via Update Instruction AX069/21/S. Letter states reason for recall, health risk and action to take: To avoid the problem, it is recommended to position the SID first and then release x-ray. Siemens will correct the software error (software VD12) via Update Instruction AX069/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. We do not consider it necessary to re-examine any patients in relation to the issues described above. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
Quantity in commerce
8 units
Distribution
US Nationwide Distribution
Product codes and lots
Serial Numbers: Serial 104007 104016 104005 104027 104012 104025 104013 132500 Expanded Recall 11/15/21: 104029 104020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0300-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0296-2022Class IITerminatedVoluntary: Firm initiated
Z-0297-2022Class IITerminatedVoluntary: Firm initiated
Z-0298-2022Class IITerminatedVoluntary: Firm initiated
Z-0299-2022Class IITerminatedVoluntary: Firm initiated
Z-0300-2022Class IITerminatedVoluntary: Firm initiated
Z-0301-2022Class IITerminatedVoluntary: Firm initiated
Z-0302-2022Class IITerminatedVoluntary: Firm initiated
Z-0303-2022Class IITerminatedVoluntary: Firm initiated
Z-0304-2022Class IITerminatedVoluntary: Firm initiated
Z-0305-2022Class IITerminatedVoluntary: Firm initiated