Z-0265-2022Class II
Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570...
Smith & Nephew, Inc. / initiated 2021-10-25 / Open, Classified / root cause: labeling and UDI
Smith & Nephew, Inc. began a recall of Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile on . FDA classified it as Class II and gives the reason as "The labels on the packages of screws were switched". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0265-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The labels on the packages of screws were switched.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- The recalling firm issued letters dated 10/25/2021 on the same day via email and FedEx overnight. The letter explained the issue and the potential risk of use of the product. Required actions were listed on the enclosed response form. The actions included inspecting their inventory and quarantining the affected product immediately and returning it. Distributors or sales representatives were to notify their customers of the field action and ensure the required actions are complete.
- Quantity in commerce
- 224 units
- Distribution
- Distribution was made to AL, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, TN, TX, and WA. There was no direct government distribution. There was no military distribution. Foreign distribution was made to Australia, Canada, China, Dubai/MENA, Germany, Italy, Japan, Netherlands, Singapore, and Thailand.
- Product codes and lots
- REF 72413575 - batch #21FM24289; REF 72413570 - batch #21GM04743.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0265-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY