Z-0482-2022Class II
Instavac Portable Suction Pump, Constant mode (model #757000, Instavac C) and Intermittent mode (model...
Ohio Medical LLC / initiated 2021-11-16 / Open, Classified / root cause: manufacturing process
Ohio Medical Corporation began a recall of Instavac Portable Suction Pump, Constant mode (model #757000, Instavac C) and Intermittent mode (model 756000, Instavac I); and Instavac replacement pumps... on . FDA classified it as Class II and gives the reason as "Increase in premature device failures". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0482-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ohio Medical LLC
- Product
- Instavac Portable Suction Pump, Constant mode (model #757000, Instavac C) and Intermittent mode (model 756000, Instavac I); and Instavac replacement pumps, part #AI5511, a component of model numbers 756000 and 757000.
- Product code
- BTA Pump, Portable, Aspiration (Manual Or Powered)
- Reason for recall
- Increase in premature device failures
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm notified their consignees by letter beginning 11/16/2021 via USPS. The letter identified the defect and requested the return of the affected devices for testing unless it is not feasible to do so. Consignees were provided an alternative option for on site testing if the device could not be returned.
- Quantity in commerce
- 649 portable suction pumps and replacement pumps
- Distribution
- US Nationwide, including PR. There was government and foreign distribution. There was no military distribution.
- Product codes and lots
- Shipped between March 2019 and February 2021.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0482-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY