Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0346-2022Class I

Puritan Bennett 980 Series Ventilator, 980X1ENDIPC

Medtronic, Inc. / initiated 2021-11-04 / Terminated / root cause: manufacturing process

Covidien Llc began a recall of Puritan Bennett 980 Series Ventilator, 980X1ENDIPC on . FDA classified it as Class I and gives the reason as "Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0346-2022
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Puritan Bennett 980 Series Ventilator, 980X1ENDIPC
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Root cause tag
manufacturing process
FDA root cause
Reprocessing Controls
Action
On November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use. Actions you should take Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies. Notify all personnel in all care environments in which the affected PB980 series ventilators are used about this medical device correction. If your facility has distributed affected PB980 series ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. Complete the attached form and return it as directed to confirm your receipt and understanding of this information. If you are aware of any incidents related to this issue, please contact our Technical Support Department immediately at 1-800-255-6774, option 4, and then option 1 to provide information regarding those events so regulatory reporting obligations can be fulfilled. Work with the Medtronic Technical Support Department if you require assistance finding alternative ventilation devices. Beginning November 5, 2021, Medtronic Technical Service will contact all affected customers to schedule a service appointment for all potentially affected ventilators. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.
Quantity in commerce
278 total
Distribution
Worldwide distribution - US Nationwide.
Product codes and lots
GTIN/UDI - 10884521172302 Serial Numbers: 35B1700447 35B1700405

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0346-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0338-2022Class ITerminatedVoluntary: Firm initiated
Z-0339-2022Class ITerminatedVoluntary: Firm initiated
Z-0340-2022Class ITerminatedVoluntary: Firm initiated
Z-0341-2022Class ITerminatedVoluntary: Firm initiated
Z-0342-2022Class ITerminatedVoluntary: Firm initiated
Z-0343-2022Class ITerminatedVoluntary: Firm initiated
Z-0344-2022Class ITerminatedVoluntary: Firm initiated
Z-0345-2022Class ITerminatedVoluntary: Firm initiated
Z-0346-2022Class ITerminatedVoluntary: Firm initiated
Z-0347-2022Class ITerminatedVoluntary: Firm initiated
Z-0348-2022Class ITerminatedVoluntary: Firm initiated
Z-0349-2022Class ITerminatedVoluntary: Firm initiated
Z-0350-2022Class ITerminatedVoluntary: Firm initiated
Z-0351-2022Class ITerminatedVoluntary: Firm initiated
Z-0352-2022Class ITerminatedVoluntary: Firm initiated
Z-0353-2022Class ITerminatedVoluntary: Firm initiated