Z-0346-2022Class I
Puritan Bennett 980 Series Ventilator, 980X1ENDIPC
Medtronic, Inc. / initiated 2021-11-04 / Terminated / root cause: manufacturing process
Covidien Llc began a recall of Puritan Bennett 980 Series Ventilator, 980X1ENDIPC on . FDA classified it as Class I and gives the reason as "Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0346-2022
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Puritan Bennett 980 Series Ventilator, 980X1ENDIPC
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Reprocessing Controls
- Action
- On November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use. Actions you should take Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies. Notify all personnel in all care environments in which the affected PB980 series ventilators are used about this medical device correction. If your facility has distributed affected PB980 series ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. Complete the attached form and return it as directed to confirm your receipt and understanding of this information. If you are aware of any incidents related to this issue, please contact our Technical Support Department immediately at 1-800-255-6774, option 4, and then option 1 to provide information regarding those events so regulatory reporting obligations can be fulfilled. Work with the Medtronic Technical Support Department if you require assistance finding alternative ventilation devices. Beginning November 5, 2021, Medtronic Technical Service will contact all affected customers to schedule a service appointment for all potentially affected ventilators. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.
- Quantity in commerce
- 278 total
- Distribution
- Worldwide distribution - US Nationwide.
- Product codes and lots
- GTIN/UDI - 10884521172302 Serial Numbers: 35B1700447 35B1700405
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0338-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0339-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0340-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0341-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0342-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0343-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0344-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0345-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0346-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0347-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0348-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0349-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0350-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0351-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0352-2022 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0353-2022 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.