iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United...

Advanced Medical Optics, Inc. / initiated 2021-10-21 / Open, Classified / root cause: software

AMO Manufacturing USA, LLC began a recall of iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer... on . FDA classified it as Class II and gives the reason as "Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0443-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Advanced Medical Optics, Inc.
Product
iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer (International) and (United States), Model: G301
Product code
LZS Excimer Laser System
Reason for recall
Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
Root cause tag
software
FDA root cause
Software design
Action
On 09/02/21, Microsoft "PrintNightmare" Vulnerability notices were sent to customers informing them of the upcoming software patches to remediate the vulnerability. This notice was posted on the recalling firm's website in the Security Advisories section. On 10/21/21 correction notices started being sent to customers. Customers were informed that Technical Service Representatives would contact then to schedule installation of the Windows OS patch. Customers were asked to do/ensure the following: 1) Complete the customer reply form, 2) To safeguard patients form potential harm and to inform them what to do in the event of product malfunction, customers were directed to specific parts of the Instructions for Use. 3) Confirm all measurements against baseline pre-operative manifest refraction and do not proceed when outside of recommended limits of difference, for sphere, cylinder and axis, as specified in the operators manual and training material. 4) Maintain manual records of patient measurements. 5) Ensure you comply with obligations under applicable privacy laws relating to putting in place safeguards in connection with personal information for patients. Customers with additional questions were encouraged to call Technical Support at 1 (800)- 511-0911.
Quantity in commerce
711
Distribution
US: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM, CA. OUS: Canada, Mexico, Russian Federation, Turkey, Brazil, Egypt, Iraq, Saudi Arabia, Jordan, Norway, Colombia, Germany, Malaysia, India, Oman, Azerbaijan, Canada, Slovakia, Lebanon, Spain, Ireland, Italy, South Korea, Bangladesh, Chile, Taiwan, Tunisia, Kazakhstan, Libya, Argentina, Singapore, Austria, Australia, Vietnam, Greece, Poland, Indonesia, Yemen, Great Britain, South Africa, Croatia, France, Czech Republic, Algeria, Thailand, Bulgaria, United Arab Emirates, Belgium, Japan, China, Nepal, Israel, Switzerland, Dominican Republic, Hong Kong, Portugal, Ecuador, Kuwait, Netherlands, Slovenia, Philippines
Product codes and lots
G300 with Windows XP. G301 with Windows 7 Embedded System.