HURRICANE RX DILATION BALLOON 8MM 2CM -for endoscopic dilatation of strictures of the biliary tree and the...
Boston Scientific Corporation began a recall of HURRICANE RX DILATION BALLOON 8MM 2CM -for endoscopic dilatation of strictures of the biliary tree and the Sphincter of Oddi. UPN: M00545930 on . FDA classified it as Class II and gives the reason as "Balloon pinhole results in losss pressure or fails to gain or maintain pressure, causing prolonged procedure". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0324-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Corporation
- Product
- HURRICANE RX DILATION BALLOON 8MM 2CM -for endoscopic dilatation of strictures of the biliary tree and the Sphincter of Oddi. UPN: M00545930
- Product code
- FGE Stents, Drains And Dilators For The Biliary Ducts
- Reason for recall
- Balloon pinhole results in losss pressure or fails to gain or maintain pressure, causing prolonged procedure
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Boston Scientific issued Urgent Medical Device Removal Letter to US consignees via Federal Express Priority Overnight mail on October 21, 2021. Letter states reason for recall, health risk and action to take: Cease further distribution or use of any remaining product affected by this removal immediately. Verify by product lot/batch number in the product table whether any product within your inventory is affected. If so, indicate on your Reply Verification Tracking Form the quantity of units from each lot/batch that you will be returning. If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately and return it to BSC in accordance with the enclosed instructions. If you are a distributor, please note that the depth of this removal is to the end user level and this notification should be forwarded to your customers. If you are a facility that has sent products to another hospital or facility within your network, please ensure that this notification is forwarded to them. Please read carefully through the removal instructions included with this notification. Your local Sales Representative can answer any questions that you may have regarding this notification.
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Batches: 25644951, 25649190, 25650286, 25651334, 25654287, 25947916, 25947918, 26010397, 26058298, 26406532, 26435581, 26449368, 26533973, 26555544, 26868194, 27074110, 27269956, 27384708. GTIN: 08714729283843 Exp. 05/27/2022 to 05/26/2023
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0320-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0321-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0322-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0323-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0324-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0325-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0326-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0327-2022 | Class II | Ongoing | Voluntary: Firm initiated |