Z-0238-2022Class II

Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)

Philips North America LLC / initiated 2021-09-24 / Open, Classified / root cause: software

Philips Healthcare began a recall of Philips Azurion systems with software releases 2.1(L1) and 2.1(L2) on . FDA classified it as Class II and gives the reason as "3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0238-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips will inform customers of the issue via an Electronic Product Radiation Correction notification. The letter identified the affected product, problem and actions to be taken. Philips Field Service Engineers will schedule an appointment with customers to install the software update. For questions contact your Philips representative at 1-800-722-9377.
Quantity in commerce
226 units
Distribution
US Nationwide distribution
Product codes and lots
Azurion 2.1(L1) and 2.1(L2)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0238-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0238-2022Class IIOngoingFDA Mandated