Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide...

Medtronic, Inc. / initiated 2021-11-12 / Open, Classified

Covidien Llc began a recall of Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and... on . FDA classified it as Class II and gives the reason as "Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0332-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA
Product code
LFL Instrument, Ultrasonic Surgical
Reason for recall
Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit
FDA root cause
Under Investigation by firm
Action
Medtronic customer communication translated by in-country representatives emailed on November 12, 2021. Letter states reason for recall, health risk and action to take: 1. Please immediately quarantine and discontinue use of the affected item code with the associated serial numbers listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated serial numbers must be returned. 3. If you have distributed the Covidien Sonicision" Battery Chargers listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. Fax to 800-895-6140 or email to [email protected] Credit for returned affected product will be issued based on the RGA number. Exception: Customers with zero inventory, fax to 651-367-2624 or email to [email protected]
Quantity in commerce
2 units OUS
Distribution
Brazil
Product codes and lots
Serial Numbers: MCSB006939, MCSB006947 GTIN: 10884521592896

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0332-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0332-2022Class IIOngoingVoluntary: Firm initiated