smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, L, 5H, 115MM, REF 72573102
Smith & Nephew, Inc. began a recall of smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, L, 5H, 115MM, REF 72573102 on . FDA classified it as Class II and gives the reason as "Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0365-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, L, 5H, 115MM, REF 72573102
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- The firm initiated the correction by email and letter on 1 NOV 2021. The notice explained the problem and requested that the consignees check their inventory and complete the Inventory Correction Certificate form. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out.
- Quantity in commerce
- 35 units
- Distribution
- US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.
- Product codes and lots
- Batch Number: 20HM10252, 20HM10253, 20HM10254, 20HM22785, UDI
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0363-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0364-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0365-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0366-2022 | Class II | Terminated | Voluntary: Firm initiated |