Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of...
Biomeme, Inc. began a recall of Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates... on . FDA classified it as Class II and gives the reason as "The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0452-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Biomeme, Inc.
- Product
- Biomeme SARS-CoV-2- Real Time RT-PCR Test, Instructions for Use, Part number 100002824 Subcomponent of: SARS-CoV-2 Go-Strips, 3000555; SARS-CoV-2 Go-Plates, 3000562; SARS-Cov-2 Bulk Vials, 3000564
- Product code
- QJR Reagents, 2019-Novel Coronavirus Nucleic Acid
- Reason for recall
- The current IFU specifies a shelf-life of one week after resuspension of the RPC Buffer when stored at room temperature. Ongoing stability studies have indicated the need to shorten the life of the RPC Buffer once resuspended from the current instruction of one week to two days when stored at room temperature (15degC-30degC).
- Root cause tag
- labeling and UDI
- FDA root cause
- Incorrect or no expiration date
- Action
- On October 20, 2021, the firm sent an informal notice to customers titled "Product Update: Storing your RNA Process Control (RPC)". On November 4, 2021, the firm followed up with a formal notice titled "Medical Device Correction". The letter notified customers of the need to shorten the life of the RPC Buffer once resuspended from one week to TWO DAYS while stored at room temperature (15 degrees Celsius - 30 degrees Celsius). Actions to be taken by the Customer: Effective immediately, users of resuspended RPC Buffer should discard after two days when stored at room temperature. For longer-term storage, the RPC Buffer can be frozen once, at a temperature of -20C, for up to 3 months, and should not be frozen and thawed more than once. The expiration date of unopened RPC Kits will remain the same. Customers were also instructed to confirm receipt of the notification through a link provided by the firm. Customers should discard any other documents they may have on-hand that specifies the longer shelf-life stated in the IFU. When available, Biomeme will deliver updated versions of the IFU to all users by email.
- Quantity in commerce
- 11,607
- Distribution
- Domestic distribution to AZ, CA, FL, HI, IN, KY, MD, MO, NC, NJ, NY, OH, OK, PA, TX, VA. Foreign distribution to Ghana, Dominican Republic, and Singapore.
- Product codes and lots
- All lots. Last lot of Kit distributed with the old IFU is 20210914.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0452-2022 | Class II | Terminated | Voluntary: Firm initiated |