LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
Covidien Llc began a recall of LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837 on . FDA classified it as Class II and gives the reason as "Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0495-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- URGENT MEDICAL DEVICE RECALL notification letters dated 12/2021 were sent to customers. Customer Action Required: 1. Please immediately quarantine and discontinue use of the affected item code with the associated lot number listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated lot number must be returned. 3. If you have distributed the product item code and lot number listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory and email it to [email protected] We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please do not hesitate to contact Azi Hosseini on +61 428 464 838, or [email protected]
- Quantity in commerce
- 504 devices
- Distribution
- Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution.
- Product codes and lots
- UDI: 20884521705880 10884521705883; Lot Number: 11320187X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0495-2022 | Class II | Terminated | Voluntary: Firm initiated |