Z-0458-2022Class II
UROMEDICA Implantation Instrument Set (Sheath and Sharp Trocar), REF 750041, nonsterile
Uromedica Inc. / initiated 2021-11-22 / Terminated / root cause: manufacturing process
Uromedica Inc. began a recall of UROMEDICA Implantation Instrument Set (Sheath and Sharp Trocar), REF 750041, nonsterile on . FDA classified it as Class II and gives the reason as "The U-channel sheaths may cause damage to ProACT implant devices when the devices are passed through the sheath during a ProACT implantation procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0458-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Uromedica Inc.
- Product
- UROMEDICA Implantation Instrument Set (Sheath and Sharp Trocar), REF 750041, nonsterile.
- Product code
- EZY Device, Incontinence, Mechanical/Hydraulic
- Reason for recall
- The U-channel sheaths may cause damage to ProACT implant devices when the devices are passed through the sheath during a ProACT implantation procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Letters dated 11/22/2021 were issued via email to the consignees explaining the reason for recall. The consignee is requested to identify and remove the affected u-channel sheath and sharp trocar from circulation at their facility. The consignee will be contacted by a Uromedica representative who will schedule an in-person visit and collect the affected U-channel sheath and sharp trocar. Once the instruments have been removed from the facility, replacement instruments will be provided.
- Quantity in commerce
- 18 devices
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, KS, MD, TX, and WI.
- Product codes and lots
- Lot numbers UM00794 and UM00817, UDI 00180668000106
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0458-2022 | Class II | Terminated | Voluntary: Firm initiated |