Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of...
Siemens Healthcare Diagnostics, Inc. began a recall of Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma Siemens... on . FDA classified it as Class II and gives the reason as "2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0440-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Atellica CH 2-Microglobulin (B2M) -M) Assay for in vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma Siemens Material Number (SMN): 11097635
- Product code
- JZG System, Test, Beta-2-Microglobulin Immunological
- Reason for recall
- 2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
- FDA root cause
- Under Investigation by firm
- Action
- Siemens issued An Urgent Medical Device Correction (UMDC) letter to US customers by FedEx beginning on 2021-12-01. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities beginning on 2021-12-01. Letter states reason for recall, health risk and action to take: Discontinue use of and discard the kit lot listed in Table 1. Review your inventory of this product to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
- Quantity in commerce
- US: 46 units OUS: 794 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CO, MA, MD, MS, NC, NE, NH, OH, VA, WI and the countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Czech Republic, Denmark, Ecuador, France, Germany, Greece, India, Iran, Israel, Italy, Japan, Mexico, Morocco, Oman, P.R. China, Pakistan, Philippines, Portugal, Slovakia, South Africa, Spain, Switzerland, Turkey, United Kingdom, Uruguay.
- Product codes and lots
- Lot Number: 211446 UDI: 00630414595566
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0440-2022 | Class II | Terminated | Voluntary: Firm initiated |