GE Vscan Extend, being sold as "Vscan Extend Sector USB"
GE Healthcare, LLC began a recall of GE Vscan Extend, being sold as "Vscan Extend Sector USB" on . FDA classified it as Class II and gives the reason as "The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0523-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- GE Vscan Extend, being sold as "Vscan Extend Sector USB"
- Product code
- IYO System, Imaging, Pulsed Echo, Ultrasonic
- Reason for recall
- The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm distributed an Urgent Medical Device Correction letter to customers on 11/17/2021 informing customers of the issue with Vscan Extend devices not powering on because of depleted internal battery. This issue occurs if the device is not turned off prior to the external battery being removed or when the device is connected to a personal computer without connecting to AC power. The actions needed to be taken by the user until a software update could be deployed were identified in this notice as: make sure to power off the Vscan Extend device before removing the external battery, always connect the device to AC power when it is connected to a PC, and if the device is being used outside of a normal hospital or clinic setting where it is the only imaging device available, ensure the device functionality is checked daily or before each shift. Attached to the notification was an Acknowledgment Response form that was to be returned by email within 30 days of receipt of the notice. Customers with any questions are to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
- Quantity in commerce
- 12,526 total reported units (61 units of this model)
- Distribution
- Domestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR, and DC. International: Albania, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dubai, Ecuador, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Haiti, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Lithuania, Malaysia, Malta, Mexico, Moldova, Morocco, Myanmar, Netherlands, New Zealand, Norway, Papua New Guinea, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Somalia, South Africa, Spain, Sweden, Switzerland, Thailand, Togo, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam, and Western Sahara.
- Product codes and lots
- Information reads "Serial Number (UDI, Model Number)". VH150315TI (01008406821216511118020021VH150315TI, H41212RR), VH15019235 (01008406821216511117033021VH15019235, H41212RR), VH150328FJ (01008406821216511118021321VH150328FJ, H41212RR), VH1503146U (01008406821216511118020921VH1503146U, H41212RR), VH1503356H (01008406821216511118021321VH1503356H, H41212RR), VH150318PI (01008406821216511118020821VH150318PI, H41212RR), VH2006357M (01008406821216511121031721VH2006357M, H41212ZB), VH1505402S (01008406821216511118120021VH1505402S, H41212ZB), VH150277MA (01008406821216511117100021VH150277MA, H41212RR), VH150278XK (01008406821216511119030221VH150278XK, H41212ZB), VH150280EN (01008406821216511117100021VH150280EN, H41212RR), VH15028191 (01008406821216511117100021VH15028191, H41212RR), VH150282CY (01008406821216511117100021VH150282CY, H41212RR), VH150283L3 (01008406821216511117100021VH150283L3, H41212RR), VH150296YP (01008406821216511118010021VH150296YP, H41212RR), VH150298DQ (01008406821216511118010021VH150298DQ, H41212RR), VH150299RZ (01008406821216511118010021VH150299RZ, H41212RR), VH15030049 (01008406821216511118010021VH15030049, H41212RR), VH150301JU (01008406821216511118010021VH150301JU, H41212RR), VH150305E7 (01008406821216511118020021VH150305E7, H41212RR), VH150325R6 (01008406821216511118020021VH150325R6, H41212RR), VH150332EL (01008406821216511118020021VH150332EL, H41212RR), VH150334AV (01008406821216511118020021VH150334AV, H41212RR), VH150356RI (01008406821216511118020021VH150356RI, H41212RR), VH150369TF (01008406821216511118020021VH150369TF, H41212RR), VH150396VW (01008406821216511118050021VH150396VW, H41212ZB), VH150398X2 (01008406821216511118050021VH150398X2, H41212ZB), VH1504078Q (01008406821216511118060021VH1504078Q, H41212ZB), VH150413SH (01008406821216511118060021VH150413SH, H41212ZB), VH1504145M (01008406821216511118060021VH1504145M, H41212ZB), VH150418YP (01008406821216511118060021VH150418YP, H41212ZB), VH15041959 (01008406821216511118110021VH15041959, H41212ZB), VH150498S4 (01008406821216511118110021VH150498S4, H41212ZB), VH1505015I (01008406821216511118110021VH1505015I, H41212ZB), VH1505026L (01008406821216511118110021VH1505026L, H41212ZB), VH150503FG (01008406821216511118110021VH150503FG, H41212ZB), VH150504PC (01008406821216511118110021VH150504PC, H41212ZB), VH150505WS (01008406821216511118110021VH150505WS, H41212ZB), VH150541JB (01008406821216511118120021VH150541JB, H41212ZB), VH150599H5 (01008406821216511119010021VH150599H5, H41212ZB), VH150603Q1 (01008406821216511119010021VH150603Q1, H41212ZB), VH1506321L (01008406821216511119030421VH1506321L, H41212ZB), VH150635LI (01008406821216511119030421VH150635LI, H41212ZB), VH150667SK (01008406821216511119040321VH150667SK, H41212ZB), VH150672TF (01008406821216511119040021VH150672TF, H41212ZB), VH150787F1 (01008406821216511119040021VH150787F1, H41212ZB), VH150788JE (01008406821216511119040021VH150788JE, H41212ZB), VH1503213I (01008406821216511118020021VH1503213I, H41212RR), VH150804V1 (01008406821216511119040021VH150804V1, H41212ZB), VH1506815J (01008406821216511119040021VH1506815J, H41212ZB), VH150409WM (01008406821216511118060021VH150409WM, H41212ZB), VH150753TC (01008406821216511119040421VH150753TC, H41212ZB), VH150421I5 (01008406821216511118080021VH150421I5, H41212ZB), VH200797WX (01008406821216511121090221VH200797WX, H41212ZB), VH150766CY (01008406821216511119050021VH150766CY, H41212ZB), VH20076196 (01008406821216511121090221VH20076196, H41212ZB), VH150520G8 (01008406821216511118120021VH150520G8, H41212ZB), VH2003248W (01008406821216511120091621VH2003248W, H41212ZB), VH150614LJ (01008406821216511119010021VH150614LJ, H41212ZB), VH1508428Q (01008406821216511119050021VH1508428Q, H41212ZB), VH200338JV (01008406821216511120091621VH200338JV, H41212ZB), VH1501821N (01008406821216511117030021VH1501821N, H41212RR), VH200729TM (01008406821216511121080521VH200729TM, H41212ZB)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0518-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0519-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0520-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0521-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0522-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0523-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0524-2022 | Class II | Terminated | Voluntary: Firm initiated |