Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300

Draeger, Inc. / initiated 2021-12-16 / Terminated / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300 on . FDA classified it as Class II and gives the reason as "26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0501-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On December 17, 2021, the firm notified customers via Urgent Medical Device Correction Notice. Customers were instructed to do the following: Please ensure that all users of the above-mentioned products and other concerned persons within your organization are made aware of this Urgent Medical Device Correction Notice. If you have provided the products to third parties, please forward a copy of this information. Please keep this information at least until the corrective measure has been completed. Please also complete and return the attached Customer Acknowledgment Card to confirm to us that you have received this information. A Drger Service representative will contact you to arrange a date for the correction to be performed free of charge. Until the repair has been concluded you may continue using the device under constant supervision as long as appropriate O2-monitoring is in place and a suitable alarm limit is set. Please consider a suitable buffer when adjusting the O2 flow. If you have any questions regarding the operation of your Fabius MRI Anesthesia Machine, please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437- 2437 (press 2 at the prompt, then 2, then 2 again). If you have any questions regarding this recall communication, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Quantity in commerce
5
Distribution
US Nationwide distribution in the states of NY, PA, and WI.
Product codes and lots
UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0501-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0501-2022Class IITerminatedVoluntary: Firm initiated