Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0511-2022Class II

KWIK-STIK 2 Pack, Catalog #0894P; KWIK-STIK 6 Pack, Catalog #0894K; and LYFO DISK, Catalog #0894L...

Microbiologics, Inc. / initiated 2020-12-16 / Terminated / root cause: manufacturing process

Microbiologics Inc began a recall of KWIK-STIK 2 Pack, Catalog #0894P; KWIK-STIK 6 Pack, Catalog #0894K; and LYFO DISK, Catalog #0894L; Microsporum canis derived from ATCC 36299 on . FDA classified it as Class II and gives the reason as "Contamination with S. epidermidis". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0511-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microbiologics, Inc.
Product
KWIK-STIK 2 Pack, Catalog #0894P; KWIK-STIK 6 Pack, Catalog #0894K; and LYFO DISK, Catalog #0894L; Microsporum canis derived from ATCC 36299.
Product code
JTR Kit, Quality Control For Culture Media
Reason for recall
Contamination with S. epidermidis
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
The recalling firm issued letters beginning 12/16/2020 to their consignees explaining the reason for recall and that due to the distinct morphology of Microsporum canis, it is possible to distinguish between the 2 colonies and select the correct microorganism from an agar plate for testing. The consignee was provided the instructions to review their lab procedures to understand how this information affects their usage, use or discard depending on their lab procedures and how this information affects their usage, complete the response form and return it to the firm. The letter is to kept for their records and the firm is to be contacted for a replacement kit if needed. The response form was to indicate the consignee has acknowledged receipt of the information, has updated their records, and to report the number of kits on hand. If the consignee has discarded or destroyed individual units, they are to indicate the quantity needed for replacement.
Quantity in commerce
59 kits
Distribution
Distribution was made to CA, IA, IL, LA, MA, MI, NY, PA, SD, TN, VA, WI, and WV. There was no government/military distribution. Foreign distribution was made to Brazil, Canada, Colombia, Dominican Republic, Finland, France, India, Japan, Malaysia, Mexico, Netherlands, Poland, Romania, Saudi Arabia, Serbia, South Africa, and United Kingdom.
Product codes and lots
KWIK-STIK 2 Pack - Catalog#: 0894P, Lot #: 894-68-3, Exp. 6/30/2022, 894-68-4, Exp. 6/30/2022, UDI: 20845357022947;  KWIK-STIK 6 Pack - Catalog#: 0894K, Lot #: 894-68-1, Exp. 6/30/2022, UDI: 30845357022951;  LYFO DISK - Catalog#: 0894L, Lot #: 894-68-2, Exp. 6/30/2022, UDI: 10845357022964;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0511-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0511-2022Class IITerminatedVoluntary: Firm initiated