Z-0511-2022Class II
KWIK-STIK 2 Pack, Catalog #0894P; KWIK-STIK 6 Pack, Catalog #0894K; and LYFO DISK, Catalog #0894L...
Microbiologics, Inc. / initiated 2020-12-16 / Terminated / root cause: manufacturing process
Microbiologics Inc began a recall of KWIK-STIK 2 Pack, Catalog #0894P; KWIK-STIK 6 Pack, Catalog #0894K; and LYFO DISK, Catalog #0894L; Microsporum canis derived from ATCC 36299 on . FDA classified it as Class II and gives the reason as "Contamination with S. epidermidis". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0511-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microbiologics, Inc.
- Product
- KWIK-STIK 2 Pack, Catalog #0894P; KWIK-STIK 6 Pack, Catalog #0894K; and LYFO DISK, Catalog #0894L; Microsporum canis derived from ATCC 36299.
- Product code
- JTR Kit, Quality Control For Culture Media
- Reason for recall
- Contamination with S. epidermidis
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- The recalling firm issued letters beginning 12/16/2020 to their consignees explaining the reason for recall and that due to the distinct morphology of Microsporum canis, it is possible to distinguish between the 2 colonies and select the correct microorganism from an agar plate for testing. The consignee was provided the instructions to review their lab procedures to understand how this information affects their usage, use or discard depending on their lab procedures and how this information affects their usage, complete the response form and return it to the firm. The letter is to kept for their records and the firm is to be contacted for a replacement kit if needed. The response form was to indicate the consignee has acknowledged receipt of the information, has updated their records, and to report the number of kits on hand. If the consignee has discarded or destroyed individual units, they are to indicate the quantity needed for replacement.
- Quantity in commerce
- 59 kits
- Distribution
- Distribution was made to CA, IA, IL, LA, MA, MI, NY, PA, SD, TN, VA, WI, and WV. There was no government/military distribution. Foreign distribution was made to Brazil, Canada, Colombia, Dominican Republic, Finland, France, India, Japan, Malaysia, Mexico, Netherlands, Poland, Romania, Saudi Arabia, Serbia, South Africa, and United Kingdom.
- Product codes and lots
- KWIK-STIK 2 Pack - Catalog#: 0894P, Lot #: 894-68-3, Exp. 6/30/2022, 894-68-4, Exp. 6/30/2022, UDI: 20845357022947; KWIK-STIK 6 Pack - Catalog#: 0894K, Lot #: 894-68-1, Exp. 6/30/2022, UDI: 30845357022951; LYFO DISK - Catalog#: 0894L, Lot #: 894-68-2, Exp. 6/30/2022, UDI: 10845357022964;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0511-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.