Z-0565-2022Class II

MIC* Percutaneous Endoscopic Gastrostomy PEG Kit - 20 Fr-PULL

Halyard Health / initiated 2021-12-22 / Terminated

Avanos Medical, Inc. began a recall of MIC* Percutaneous Endoscopic Gastrostomy PEG Kit - 20 Fr-PULL on . FDA classified it as Class II and gives the reason as "The External Retention Bolster contained in the kit may be a larger diameter than the tube". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0565-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
MIC* Percutaneous Endoscopic Gastrostomy PEG Kit - 20 Fr-PULL
Product code
KNT Tubes, Gastrointestinal (And Accessories)
Reason for recall
The External Retention Bolster contained in the kit may be a larger diameter than the tube.
FDA root cause
Under Investigation by firm
Action
All USA Distributors and healthcare facilities that were directly shipped impacted devices were sent a Product Advisory Notice beginning on December 22, 2021 by email. Clinical Consignees were instructed to evaluate their facilities inventory of products to determine if any of the impacted product codes were present in their facility. The letter provided information identifying the affected product. It was indicated to segregate and return impacted product. The clinician may recognize that the External Retention Bolster is too large for the accompanying PEG tube when initially use the product. Consignees were instructed to complete an Acknowledgement Form and return a copy of the form to Avanos within five (5) days.
Quantity in commerce
2652 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and Chile.
Product codes and lots
1) Code 7160-20, Lots 30116959, 30132543, 30132545, and 30132546 2) Code 7640-20, Lots 30129345

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0565-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0565-2022Class IITerminatedVoluntary: Firm initiated