Z-0566-2022Class II
MIC* Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors - 20 Fr-PULL
Halyard Health / initiated 2021-12-22 / Terminated
Avanos Medical, Inc. began a recall of MIC* Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors - 20 Fr-PULL on . FDA classified it as Class II and gives the reason as "The External Retention Bolster contained in the kit may be a larger diameter than the tube". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0566-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Halyard Health
- Product
- MIC* Percutaneous Endoscopic Gastrostomy PEG Kit with ENFit Connectors - 20 Fr-PULL
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- The External Retention Bolster contained in the kit may be a larger diameter than the tube.
- FDA root cause
- Under Investigation by firm
- Action
- All USA Distributors and healthcare facilities that were directly shipped impacted devices were sent a Product Advisory Notice beginning on December 22, 2021 by email. Clinical Consignees were instructed to evaluate their facilities inventory of products to determine if any of the impacted product codes were present in their facility. The letter provided information identifying the affected product. It was indicated to segregate and return impacted product. The clinician may recognize that the External Retention Bolster is too large for the accompanying PEG tube when initially use the product. Consignees were instructed to complete an Acknowledgement Form and return a copy of the form to Avanos within five (5) days.
- Quantity in commerce
- 692 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada and Chile.
- Product codes and lots
- 1) Code 8160-20, Lot 30135103 2) Code 8640-20, Lot 30132544
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0565-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0566-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.