Z-0479-2022Class II

Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032

PTW North America Corporation / initiated 2021-08-27 / Terminated / root cause: software

PTW NORTH AMERICA CORPORATION began a recall of Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032 on . FDA classified it as Class II and gives the reason as "When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0479-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
PTW North America Corporation
Product
Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032
Product code
IYE Accelerator, Linear, Medical
Reason for recall
When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results
Root cause tag
software
FDA root cause
Software design
Action
PTW US notified customer Urgent Safety Notice via email and telephone on 8/27/20. Letter states reason for recall, health risk and action to take: PTW will provide free-of-charge software updates. Users who use BeamAdjust 2.2/ VeriSoft 8.0 according to the above described application must update the software to VeriSoft 8.0.1/BeamAdjust 2.2.1 prior to further use of the software. For all other users PTW recommends using BeamAdjust 2.2.1/VeriSoft 8.0.1 too. Until the software updates are available PTW recommends avoiding the air density correction method. As a workaround, if air density correction should still be applied a manually calculated air density correction value can be entered as a kUser keeping kTP untouched (kTP =1.000 !). Please also note, machine QA measurements performed with PTW detector arrays within the PTW software MultiCheck are not affected. TW has fixed the bug in the software and provides all affected users with software updates.
Quantity in commerce
1 unit
Distribution
US Nationwide distribution in the state of CO.
Product codes and lots
S/N :AU2001094-1 UDI: EPTWS0700090 VeriSoft ;EPTWS0700090 BeamAdjust

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0479-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0479-2022Class IITerminatedVoluntary: Firm initiated