Z-0512-2022Class II

Neodent Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 -intended for dental implant Item 138.089

Straumann USA, LLC / initiated 2021-11-30 / Terminated / root cause: packaging

Straumann USA LLC began a recall of Neodent Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 -intended for dental implant Item 138.089 on . FDA classified it as Class II and gives the reason as "Mislabeled: Carton box item 138.089- Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 has an implant with measure of 3.5x13mm; and Carton Box (138.107- Easypack Helix GM Acqua has a 3.5x8mm implant instead of a 3.5x 13mm...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0512-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Straumann USA, LLC
Product
Neodent Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 -intended for dental implant Item 138.089
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
Mislabeled: Carton box item 138.089- Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 has an implant with measure of 3.5x13mm; and Carton Box (138.107- Easypack Helix GM Acqua has a 3.5x8mm implant instead of a 3.5x 13mm implant
Root cause tag
packaging
FDA root cause
Packaging change control
Action
Neodent issued Medical Device Recall letter initial contact by phone initiated on November 30th, 2021; followed by Confirmation written letter via UPS December 3rd 2021. Letter states reason for recall, health risk and action to take: Search in your stock for units according to the description on the table above. " If the implant has not yet been used, identify and segregate the affected units, open or closed, in your stock. Return the product using the label included in this mailing, refer to Field Action . Upon receipt of the product by Straumann, replacement product will be sent. If the product is not found in your inventory, indicate on the Form below. " If the product has been implanted and no issue was identified, it is not necessary to remove, additional patient follow-up is not required. Indicate this on the Customer Confirmation Form (included) " If the product has been installed and an issue has been reported by the patient, for example: numbness, pain or any other symptom, please communicate immediately to Straumann. " For either cases, complete and return the enclosed Customer Confirmation Form to Action 003/2021. 3.1. DATA FOR PRODUCT RETURN: Straumann USA Regulatory Affairs Action 003/2021 Product Return 60 Minuteman Road, Andover, MA 01810
Quantity in commerce
7 pieces
Distribution
US Nationwide Distribution in the states of FL, IL, OH
Product codes and lots
Lot GNV17 UDI: 7899878052722

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0512-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0512-2022Class IITerminatedVoluntary: Firm initiated
Z-0513-2022Class IITerminatedVoluntary: Firm initiated