Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0550-2022Class I

OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of...

Abiomed, Inc. / initiated 2021-12-14 / Terminated / root cause: software

Abiomed, Inc. began a recall of OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The... on . FDA classified it as Class I and gives the reason as "The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0550-2022
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abiomed, Inc.
Product
OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.
Product code
DTZ Oxygenator, Cardiopulmonary Bypass
Reason for recall
The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
The firm began notifying customers by phone starting December 14, 2021. The firm followed up by sending Urgent Medical Device Recall letters to customers on January 20, 2022. The firm is requesting that customers return their devices. The firm does not recommend placing new patients on support. For patients currently on support with an affected system, the firm recommends physicians use their clinical judgement on whether it is appropriate to continue a patient on the OXY-1 System or to transition that patient to the next step in their clinical journey. The letter included instructions on how to return impacted product. If you have questions on returning product, If you have any questions, please contact Diane Welsh from Abiomed s customer service team at (978) 646-1467 or email [email protected]. The firm's Clinical field staff will be available to coordinate boxing and shipping of the OXY-1 system for return to Abiomed. If you have questions regarding this recall, please contact Ralph Barisano directly at (978) 882-8068.
Quantity in commerce
42 Units
Distribution
US distribution to FL, GA, IN, MD, NC, PA, TX.
Product codes and lots
UDI 00860001797909 Software Version Number V-1.3.15.0, V-1.4.6.0 and V.1.4.8.0 Serial Numbers 20200231-04 20200249-01 20200249-05 20200309-02 20200339-01 20200339-03 20200352-01 20200442-01 20200442-02 20200442-05 20210768-01 20210795-02 20210870-01 20210880-01 20210890-01 20210894-01 20210895-01 20210896-01 20210897-01 20210795-04

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0550-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0550-2022Class ITerminatedVoluntary: Firm initiated