Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm...
LivaNova USA, Inc. began a recall of Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model 303-30... on . FDA classified it as Class II and gives the reason as "A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0532-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model 303-30 PereniaDURA Lead (3.0 mm electrode inner diameter)
- Product code
- LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
- Reason for recall
- A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- On 12/02/21, Liva Nova initiated a verbal notification to hospital consignees. Hospitals were asked to remove affected devices from inventory. On 12/17/21, a removal notice was distributed to consignees who were asked to take the following actions: 1. Remove affected devices from your hospital/facility s inventory. 2. Recalling firm will contact you to coordinate removal and replacement of these devices. 4. If affected devices have been implanted, please notify your local representative as soon as possible. A medical device correction notice was sent to doctors managing patients with implanted affected devices were asked to take the following actions: 1. Ask the patient to report any changes in the comfort or perception of their stimulation therapy. 2. Advise patients to use their magnet daily (epilepsy indication only) to verify that stimulation is felt, and to report any change in perceived clinical symptoms related to stimulation. An inability to feel stimulation or changes in clinical symptoms may be an early indicator of a lead break (e.g., increase in seizures, painful stimulation, or changes in the perception of stimulation might indicate a lead break). 3. At each office visit, including those scheduled due to changes reported by patient for items 1 or 2 above, interrogate the device and run device system diagnostics . If diagnostics suggest that a lead fracture is present, program the VNS generator to 0mA Please ensure this notice is communicated to all personnel within your organization who need to be aware of it. This action is being reported to the Food and Drug Administration. For questions customers can contact the recalling firm at: +1 (281) 228-7330 (Monday to Friday, 8 AM to 5 PM CST) or e-mail at [email protected].
- Quantity in commerce
- 10
- Distribution
- US: NE, MO, AR, WV, KS, MA, IN
- Product codes and lots
- UDI: 05425025750115. Serial Numbers: 119441, 119442, 119443, 119452, 119453, 119454, 119455, 119456, 119457, 119458
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0532-2022 | Class II | Terminated | Voluntary: Firm initiated |