Z-0533-2022Class II

X060-0270, 8mm Titanium Insert, Axle Interspinous Fusion System

Xtant Medical Holdings, Inc. / initiated 2021-12-29 / Terminated

XTANT Medical began a recall of X060-0270, 8mm Titanium Insert, Axle Interspinous Fusion System on . FDA classified it as Class II and gives the reason as "The proximal opening of the inserts exhibit an out of specification condition which would prevent the crossbar plates from appropriately engaging with the insert". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0533-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Xtant Medical Holdings, Inc.
Product
X060-0270, 8mm Titanium Insert, Axle Interspinous Fusion System.
Product code
PEK Spinous Process Plate
Reason for recall
The proximal opening of the inserts exhibit an out of specification condition which would prevent the crossbar plates from appropriately engaging with the insert.
FDA root cause
Under Investigation by firm
Action
On 12/30/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via FedEx Priority Overnight (Direct Signature Required) to its distributes to notify their customer that a specific lot of their interspinous fusion systems were manufactured out of tolerance resulting in the distal end of the crossbar not properly engaging with the inserts. Customer are instructed to: 1. Immediately examine their inventory and quarantine any affected products subject to this recall. 2. Complete the enclosed Acknowledgement Form, even if customers have no remaining devices in their inventory, and return the completed Acknowledgment Form to the attention of Michele Dolan via: Mail: Xtant Medical Axle Interspinous Fusion System Recall 664 Cruiser Lane Belgrade, MT 59714 Email: [email protected] Fax: 1-406-388-3380 3. Return Affected Product Contact Michele Dolan at [email protected], or 1-406-388-0480 ext. 1127, between 8 a.m. and 5 p.m. (Mountain Time) to obtain instructions for returning affected product, and to obtain replacement devices, as applicable, which will be provided at no cost to you. 4. Package the affected products to be returned. Xtant will provide a shipping container (if needed) with a return shipping label. Securely package the affected product(s). Send the container at your earliest opportunity to: ATTN: Axle Interspinous Fusion Recall RA# 21-xxxx Xtant Medical 732 Cruiser Lane Belgrade, MT 59714 5. To Inform all their staff members of this recall and if affected products have been further distributed to contact all customers. For any questions contact the Director of Regulatory Affairs/Quality Assurance at 1-406-388-0480 ext. 1125 from 8 a.m. to 5 p.m. (Mountain Time).
Quantity in commerce
24 devices
Distribution
U.S.: CA, MI, NV, and TX O.U.S.: Portugal
Product codes and lots
X060-0270, 8mm Titanium Insert, Axle System UDI (Primary DI Number): M697X06002701/$$7058963W Lot Number: 058963

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0533-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0533-2022Class IITerminatedVoluntary: Firm initiated