Z-0548-2022Class II
INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND
Smith & Nephew, Inc. / initiated 2022-01-04 / Open, Classified / root cause: manufacturing process
Smith & Nephew, Inc. began a recall of INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND on . FDA classified it as Class II and gives the reason as "The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0548-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Smith & Nephew, Inc.
- Product
- INTEGRA SURFIX ALPHA Screwdriver Torx 10, REF 219 135 ND
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated the recall by letter and email on 01/04/2022. The letter requested that the consignee examine their inventory, quarantine all affected units, and return the completed response form. Entities who have further distributed the product were directed to notify their customers. The firm is seeking return of the product.
- Quantity in commerce
- 8 units (US)
- Distribution
- US Nationwide distribution in the states of CA and OH.
- Product codes and lots
- UDI: (01)10381780052234 (11)21101 (10)FVRG, Batch Numbers: FTL9, FTLA, and FVRG