KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM)...
Microbiologics Inc began a recall of KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM), Lactobacillus acidophilus derived from ATCC(R) 4356(TM) on . FDA classified it as Class III and gives the reason as "The expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0546-2022
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microbiologics, Inc.
- Product
- KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM), Lactobacillus acidophilus derived from ATCC(R) 4356(TM)
- Product code
- JTR Kit, Quality Control For Culture Media
- Reason for recall
- The expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should have been 7/31/2022.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Microbiologics notified customers on about 12/16/2020 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions were to update records with the correct expiration date of 07/31/2022 and keep the letter for their records. Questions or concerns can be directed to Recall Support team at 320.229.7073 or [email protected].
- Distribution
- US Nationwide distribution in the states of AL, CA, CO, IL, MD, MO, MS, NM, OH, OR, PA, TX, VT.
- Product codes and lots
- Main Product Codes: Catalog Number: 8164, Lot Number: 8164-78, UDI: 70845357028623, Expiration date: 09/30/2021. Affected Component Codes: Catalog Number: 0243U, Lot Number: 243-48-5, UDI: 10845357004083, Expiration Date on Label: 07/31/2020, Correct Expiration Date: 07/31/2022.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0546-2022 | Class III | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.