KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM)...

Microbiologics, Inc. / initiated 2020-12-16 / Terminated / root cause: manufacturing process

Microbiologics Inc began a recall of KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM), Lactobacillus acidophilus derived from ATCC(R) 4356(TM) on . FDA classified it as Class III and gives the reason as "The expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should have...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0546-2022
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microbiologics, Inc.
Product
KWIK-STI(TM), GBS QC Set (Live Culture), REF 0243U. Specifically affected component: KWIK-STIK(TM), Lactobacillus acidophilus derived from ATCC(R) 4356(TM)
Product code
JTR Kit, Quality Control For Culture Media
Reason for recall
The expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should have been 7/31/2022.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Microbiologics notified customers on about 12/16/2020 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions were to update records with the correct expiration date of 07/31/2022 and keep the letter for their records. Questions or concerns can be directed to Recall Support team at 320.229.7073 or [email protected].
Distribution
US Nationwide distribution in the states of AL, CA, CO, IL, MD, MO, MS, NM, OH, OR, PA, TX, VT.
Product codes and lots
Main Product Codes: Catalog Number: 8164, Lot Number: 8164-78, UDI: 70845357028623, Expiration date: 09/30/2021. Affected Component Codes: Catalog Number: 0243U, Lot Number: 243-48-5, UDI: 10845357004083, Expiration Date on Label: 07/31/2020, Correct Expiration Date: 07/31/2022.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0546-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0546-2022Class IIITerminatedVoluntary: Firm initiated