Z-0725-2022Class II

Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an...

Philips North America LLC / initiated 2021-11-24 / Open, Classified / root cause: manufacturing process

Philips North America Llc began a recall of Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031 on . FDA classified it as Class II and gives the reason as "Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0725-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Philips Issued Medical Device Correction letter dated 11/24/21 to U.S.A, Customers via certified mailing from the United States Postal Service (USPS). Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk and action to take: Philips will inspect the affected systems for leaks and repair as required. In addition, Philips will install an extended drip tray to prevent damage from any future leaks. This service will be provided free of charge for affected systems. A Philips service representative will contact customers with affected devices to arrange for the service. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377 (reference FCC 72200484).
Quantity in commerce
1 unit
Distribution
US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippines Russian Federation Saudi Arabia South Africa Spain Sri Lanka Switzerland Taiwan Thailand Turkey United Kingdom Virgin Islands (U.S.)
Product codes and lots
S/N: OUS 109

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0725-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0724-2022Class IIOngoingVoluntary: Firm initiated
Z-0725-2022Class IIOngoingVoluntary: Firm initiated