Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0510-2022Class II

Voyager Rollator-intended as a mechanical walker Model Number: RLEU10xx (RLEU10BL, RLEU10PK, RLEU10WT)...

Compass Health Brands (Corporate Office) / initiated 2022-01-10 / Open, Classified / root cause: design

Compass Health Brands (Corporate Office) began a recall of Voyager Rollator-intended as a mechanical walker Model Number: RLEU10xx (RLEU10BL, RLEU10PK, RLEU10WT) RLEU10BL Voyager Rollator Cobalt Blue UPC: 815067078066... on . FDA classified it as Class II and gives the reason as "Wheel spoke may crack causing the wheel to separate from the axle may result in patient injuries, including bruising and lacerations requiring stitches". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0510-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Compass Health Brands (Corporate Office)
Product
Voyager Rollator-intended as a mechanical walker Model Number: RLEU10xx (RLEU10BL, RLEU10PK, RLEU10WT) RLEU10BL Voyager Rollator Cobalt Blue UPC: 815067078066; RLEU10PK Voyager Rollator Rose Gold UPC: 815067078073; RLEU10WT Voyager Rollator Ice Palace UPC: 815067078080
Product code
ITJ Walker, Mechanical
Reason for recall
Wheel spoke may crack causing the wheel to separate from the axle may result in patient injuries, including bruising and lacerations requiring stitches.
Root cause tag
design
FDA root cause
Device Design
Action
Compass Health issued Urgent Medical Device recall letter dated January 2022 on 1/11/22 issued to Distributors and Customers via email or hard copy letter. Firm posted recall to their website: https://compasshealthbrands.com/product-recall . Letter states reason for recall, health risk and action to take: Users: Users should stop using the subject products and contact Compass Health Brands. REQUIRED ACTION: 1. The Voyager Rollator casters and axles require replacement. 2. Contact Compass Health Brands at (800) 947 1728 to order replacement kits, P/N Voyager Kit. This kit contains all the parts necessary to complete replacement of the Voyager Rollator casters and axles. 3. A Voyager Caster Installment Guide will be sent as part of the Field Action Kit P/N Voyager Kit and are posted on the Compass Health Brands website. No special certification is required. 4. Complete and return the attached Consumer Field Correction Response Form to Compass Health Brands within fifteen (15) days of receipt of this field correction notification confirming your acknowledgement. Send completed form to Compass Health Brands via email at [email protected]. Complete and return this form even if you do not have affected product on hand. 5. Field correction repairs must be completed on affected units before further use of the unit. Distributors: The manufacturer has changed the wheel spoke design to increase the spoke thickness and strength. The manufacturer will supply replacement wheels with the improvement. The improved wheels will be made available to users to replace the existing wheels on the product. 1. Ensure all affected personnel are fully informed of this notice. 2. Forward this notice to your Regulatory Compliance Manager, Purchasing Manager, Customer Service Manager and Field Technicians. 3. The Voyager Rollator casters and axles require replacement. 4. Dealers/distributors or end users should contact Compass Health Brands at (800) 947-1728 and to order replacem
Quantity in commerce
3734 units
Distribution
Nationwide
Product codes and lots
Lot/Serial # RM2005000001-RM21004005000 (May 2020 through April 2021 manufacture dates)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0510-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0510-2022Class IIOngoingVoluntary: Firm initiated