Z-0588-2022Class II

QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In...

Microbiologics, Inc. / initiated 2020-02-18 / Terminated / root cause: manufacturing process

Microbiologics Inc began a recall of QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms on . FDA classified it as Class II and gives the reason as "Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0588-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microbiologics, Inc.
Product
QC Sets and Panels, KWIK-STIK AST-GP (6 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms
Product code
JTR Kit, Quality Control For Culture Media
Reason for recall
Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Microbiologics notified customers on about 02/18/2020 via letter titled "Urgent Medical Device Recall." Instructions included to inspect inventory for affected products. If found, instructions included the following: 1. REVIEW lab procedures to understand how thisinformation affects your usage; 2. USE OR DISCARD depending on your lab procedures and how this information affects your usage; 3. COMPLETE the response form provided; 4. RETURN the response form to [email protected]; 5. KEEP this letter for your records; 6. CONTACT Microbiologics if a replacement kit is needed. Questions or concerns can be directed to Technical Support team at 320.229.7045 or [email protected].
Quantity in commerce
139 units
Distribution
Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Product codes and lots
Catalog Number: 5220P; Lot Numbers (Expiration Date): 5220-01 (04/30/2020), 5220-02 (07/31/2020), 5220-03 (10/31/2020), 5220-04 (10/31/2020), 5220-05 (10/31/2020), 5220-06 (01/31/2021); UDI: 70845357026551

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0588-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0585-2022Class IITerminatedVoluntary: Firm initiated
Z-0586-2022Class IITerminatedVoluntary: Firm initiated
Z-0587-2022Class IITerminatedVoluntary: Firm initiated
Z-0588-2022Class IITerminatedVoluntary: Firm initiated
Z-0589-2022Class IITerminatedVoluntary: Firm initiated