Z-0722-2022Class II
Thermo Scientific Sensititre Gram Negative Susceptibility Testing Plate, IVD format
Remel Inc / initiated 2022-01-10 / Open, Classified
Remel Inc began a recall of Thermo Scientific Sensititre Gram Negative Susceptibility Testing Plate, IVD format on . FDA classified it as Class II and gives the reason as "There are lower than expected MICs for some gram negative species". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0722-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Remel Inc
- Product
- Thermo Scientific Sensititre Gram Negative Susceptibility Testing Plate, IVD format
- Product code
- JWY Manual Antimicrobial Susceptibility Test Systems
- Reason for recall
- There are lower than expected MICs for some gram negative species.
- FDA root cause
- Under Investigation by firm
- Action
- A customer letter was issued to impacted consignees on or around 01/10/2022. Customers are asked to check for affected product and the recalling firm requests that customers do not report results of cefiderocol when affected product is used with Thermo Scientific Mueller-Hinton Broth with TES, catalong T3462, lots 267261 to 402775. Consignees are asked to disseminate the recall information within their organization and to any other organization to which affected products have been transferred. An acknowledgement form should be submitted by email to [email protected].
- Quantity in commerce
- 1210 devices
- Distribution
- Distribution throughout the United States.
- Product codes and lots
- Product Number MDRGN2F; Lots: 1) B1402A (Expiration 2022/Oct05) 2) B1315A (Expiration 2022/Aug06) 3) B1211 (Expiration 2022/May24) 4) B1161A (Expiration 2022/Apr19 5) B0361 (Expiration 2021/Aug31) 6) B0164B (Expiration 2021/Apr16)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0722-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY