Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0661-2022Class II

IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software

Natus Medical Incorporated / initiated 2021-12-02 / Terminated

Natus Neurology DBA Excel Tech., Ltd. (XLTEK) began a recall of IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software on . FDA classified it as Class II and gives the reason as "Tile kits that allow one to install and situate a video camera within a patient environment are not fitting properly in some domestic institution ceiling tile grids that have a thinner style grid frame. The ceiling tile...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0661-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Natus Medical Incorporated
Product
IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software
Product code
OMB Automatic Event Detection Software For Full-Montage Electroencephalograph
Reason for recall
Tile kits that allow one to install and situate a video camera within a patient environment are not fitting properly in some domestic institution ceiling tile grids that have a thinner style grid frame. The ceiling tile may suddenly fall from the ceiling and cause a safety risk to the patient or end user.
FDA root cause
Process change control
Action
On 12/02/21, correction notices were mailed to customers informing them to stop using this device immediately until a solution to this issue is found. On 12/08/21, correction notices were mailed to customers informing them to follow the steps in document DOC-057825 IPELA FHD PTZ TILE KITS Correction to install Safety Wires on ceilings that meet the following structural requirements: 1) Ceilings must have existing 2' x 2' (609.6 mm x 609.6mm) ceiling tile. 2) Tile Kits ceiling tile is only compatible with a 15/16" (23.8 mm) grid tee. 3) The suspended ceiling grid must be able to support the full weight of the TILE KITS ceiling tile of 19 lbs (8.6 kg) without exceeding its loading capacity. 4) Install safety wires following instructions. Customers were asked to complete the customer reply form and to email it to [email protected] Customers with questions were provided the following phone number: 800-387-7516 Customers were asked to pass the correction notice to all those who need to be aware within their organization or to any organization where the potentially affected devices may have been transferred.
Quantity in commerce
221
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, IL, TX, MI, MA, NE, OK, WV, MO, VA, CT, IN, HI, PA, CA, MD, OH, CO, SC, PR, NV, NJ, OR, LA, DC, NC and the countries of Nepal, Viet Nam, Canada, Saudi Arabia, Indonesia, Germany, Korea, Republic of
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0661-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0661-2022Class IITerminatedVoluntary: Firm initiated