Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0634-2022Class II

Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available...

Synthes (USA) Products LLC / initiated 2021-12-22 / Open, Classified / root cause: design

Synthes (USA) Products LLC began a recall of Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System... on . FDA classified it as Class II and gives the reason as "Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0634-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Synthes (USA) Products LLC
Product
Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.
Root cause tag
design
FDA root cause
Device Design
Action
Depuy Synthes issued Urgent Medical Device Recall (Removal) issued on 12/22/21 via email FedX. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have the subject products and quarantine the subject product. DO NOT USE THE SUBJECT PRODUCT. 2. Contact your DePuy Synthes Sales Consultant or call the customer support services at 1-800-479-6329 to coordinate the return/credits of the subject product. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to [email protected] within 3 business days of receipt of this notification. Please include in the email subject: FA 2050712: 2021 Universal Chuck. " Please complete the attached Business Response Form even if you do not have the subject product on hand. 4. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 5. If any of the subject product has been forwarded to another facility, contact that facility and provide them with this notice. 6. Post a copy of this notice in a visible area for awareness and keep a copy for your records. Should you have any inquiries please do not hesitate to contact your DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir
Quantity in commerce
600 units
Distribution
US Nationwide distribution.
Product codes and lots
Lot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0634-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0634-2022Class IIOngoingVoluntary: Firm initiated