Z-0647-2022Class II

Zenition 50, Model #718096

Philips Medical Systems Nederland B.V. / initiated 2021-12-02 / Open, Classified / root cause: software

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Zenition 50, Model #718096 on . FDA classified it as Class II and gives the reason as "Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0647-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Zenition 50, Model #718096
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
On December 2, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. The Medical Device Correction includes the following instructions for the Customer: o Ensure that the wired foot switch is always connected to the system. o Immediately start using the wired foot switch or hand switch in the event the wireless foot switch loses connection. o Perform a daily check for damage and correct functioning of the hand switch and wired foot switch as described by the User Routine Checks section in your Instruction for Use. o Circulate this notice to all users of this device so that they are aware of the issue./ o Place this Medical Device Correction with the documentation of the Philips Zenition 50, Zenition 70 or Veradius Unity system. o Return the attached reply form to Philips to confirm that: 1. The Zenition 50, Zenition 70 and Veradius Unity systems wired foot switch is connected to the system 2. The users of the system have reviewed and understood this Medical Device Correction. If you need further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377.
Quantity in commerce
16 US; 76 ROW
Distribution
Domestic distribution to AZ CA FL GA HI IL KS KY MA MD ME MI MN MS NC NE NJ NV NY OH PA TN TX VA WA WV. Worldwide foreign distribution.
Product codes and lots
Wireless Foot switch 3P : 459801238191  Wireless Foot switch 3P: 459801238231

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0647-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0647-2022Class IIOngoingVoluntary: Firm initiated
Z-0648-2022Class IIOngoingVoluntary: Firm initiated
Z-0649-2022Class IIOngoingVoluntary: Firm initiated