Z-0712-2022Class II

KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived...

Microbiologics, Inc. / initiated 2014-10-06 / Terminated / root cause: manufacturing process

Microbiologics Inc began a recall of KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus... on . FDA classified it as Class II and gives the reason as "This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0712-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microbiologics, Inc.
Product
KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299
Product code
JTR Kit, Quality Control For Culture Media
Reason for recall
This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Microbiologics notified customers on about 10/06/2014 via letter titled "E.faecalis Gentamicin Susceptibility Notification." Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall, it is recommended that the intended use of the organism be reviewed to determine if Gentamicin synergy results have an impact on use. It is recommended to discard these lots if an impact is to be expected. Questions or concerns can be directed to the Technical Support team at 320.229.7064 or [email protected].
Quantity in commerce
1,873 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, United Arab Emirates, and Vietnam.
Product codes and lots
a. Catalog Number 0959P, UDI: 20845357023210, Lot Number (Expiration Date): 959-74-2 (10/08/2014), 959-74-3 (10/08/2014), 959-74-7 (10/08/2014), 959-75-4 (04/22/2015), 959-75-8 (04/22/2015), 959-75-9 (04/22/2015), 959-75-14 (04/22/2015), 959-76-1 (07/08/2015), 959-77-3 (10/21/2015); b. Catalog Number 0959K, UDI: 30845357023224, Lot Number (Expiration Date): 959-74-4 (10/08/2014), 959-74-6 (10/08/2014), 959-74-8 (10/08/2014), 959-75-1 (04/22/2015), 959-75-8 (04/22/2015), 959-75-12 (04/22/2015), and 959-77-1 (10/21/2015)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0712-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0712-2022Class IITerminatedVoluntary: Firm initiated
Z-0713-2022Class IITerminatedVoluntary: Firm initiated
Z-0714-2022Class IITerminatedVoluntary: Firm initiated