LYFO-DISK, Quality control kit for culture media, sold as: LYFO-DISK Enterococcus faecalis derived from ATCC...
Microbiologics Inc began a recall of LYFO-DISK, Quality control kit for culture media, sold as: LYFO-DISK Enterococcus faecalis derived from ATCC 51299 on . FDA classified it as Class II and gives the reason as "This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0713-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microbiologics, Inc.
- Product
- LYFO-DISK, Quality control kit for culture media, sold as: LYFO-DISK Enterococcus faecalis derived from ATCC 51299
- Product code
- JTR Kit, Quality Control For Culture Media
- Reason for recall
- This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Microbiologics notified customers on about 10/06/2014 via letter titled "E.faecalis Gentamicin Susceptibility Notification." Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall, it is recommended that the intended use of the organism be reviewed to determine if Gentamicin synergy results have an impact on use. It is recommended to discard these lots if an impact is to be expected. Questions or concerns can be directed to the Technical Support team at 320.229.7064 or [email protected].
- Quantity in commerce
- 153 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, United Arab Emirates, and Vietnam.
- Product codes and lots
- Catalog Number 0959L, UDI: 10845357023237, Lot Number (Expiration Date): 959-74-5 (10/08/2014), 959-74-8 (10/08/2014), 959-75-1 (04/22/2015), 959-75-5 (04/22/2015), 959-75-11 (04/22/2015), and 959-77-2 (10/21/2015)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0712-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0713-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0714-2022 | Class II | Terminated | Voluntary: Firm initiated |