Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0693-2022Class II

Regard, TOTAL HIP - SPRINGFIELD, Item No. 8800297010, Case Quantity: 2, Sterile

ROi CPS LLC / initiated 2022-01-13 / Terminated / root cause: labeling and UDI

ROi CPS LLC began a recall of Regard, TOTAL HIP - SPRINGFIELD, Item No. 8800297010, Case Quantity: 2, Sterile on . FDA classified it as Class II and gives the reason as "Products are mislabeled with the incorrect product identification on individual kit labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0693-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ROi CPS LLC
Product
Regard, TOTAL HIP - SPRINGFIELD, Item No. 8800297010, Case Quantity: 2, Sterile
Product code
OJH Orthopedic Tray
Reason for recall
Products are mislabeled with the incorrect product identification on individual kit labeling.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
The firm emailed a recall notification to customers on 01/13/2022 tited "ROi CPS, LLC Recall-Custom Procedure Kit Recall-Kits Containing the Incorrect Product Label." The recall notice states that the reason for recall is due to the individual kits being incorrectly labeled, while the outer box is labeled correctly. The outer box was labeled with Product No. 880297010, while the inner kit bags were labeled as Product No. 880261009. Customers are directed to review their inventory to determine if recalled kits are in stock and quarantine any identified recalled product. The firm instructs customers to open the outer carton to examine the packaging of individual bags for incorrect labeling; do not open sterile barrier packaging. Cases containing the correct inner packaging can be released. Customers are asked to complete the recall response form attached to the recall notification indicating any affected product on hand. Complete forms are to be emailed back to [email protected] so product return can be arranged.
Quantity in commerce
50 kits
Distribution
Missouri
Product codes and lots
Item No. 8800297010; Lot No. 90328; UDI 10194717113164

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0693-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0693-2022Class IITerminatedVoluntary: Firm initiated