Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0669-2022Class II

RESODURA matrix onlay 10.0 cm x 12.5 cm - indicated as dural substitutes for the repair of dura matter. REF...

Collagen Matrix, Inc / initiated 2021-12-13 / Terminated / root cause: manufacturing process

Collagen Matrix, Inc. began a recall of RESODURA matrix onlay 10.0 cm x 12.5 cm - indicated as dural substitutes for the repair of dura matter. REF: DUMO10125 on . FDA classified it as Class II and gives the reason as "Outer packaging (outer pouch) is not sealed compromising sterility and result in potential risk of patient infection, which could lead to revision surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0669-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Collagen Matrix, Inc
Product
RESODURA matrix onlay 10.0 cm x 12.5 cm - indicated as dural substitutes for the repair of dura matter. REF: DUMO10125
Product code
GXQ Dura Substitute
Reason for recall
Outer packaging (outer pouch) is not sealed compromising sterility and result in potential risk of patient infection, which could lead to revision surgery
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On Monday, December 6, 2021, Collagen Matrix Inc began informing the distributors (initial consignees) to place the lot in question on ship hold. On Monday, December 13, 2021, Collagen Matrix began issuing Urgent Medical Device Recall Notice letter to distributors via email. Letter states reason for recall, health risk and action to take: Check your inventory and that of your sub-distributors if applicable to determine if any of the affected product remains in stock and segregate all affected units from your inventory for return. Complete the Certificate of Acknowledgement form with the appropriate information and return within 5 days of receiving this notice. The completed form can be emailed to [email protected] or returned by mail to: Product Recall Collagen Matrix, Inc. 15 Thornton Road Oakland, NJ 07436 Upon receipt of the completed form, Collagen Matrix will ship replacement product and a prepaid return shipping label will be provided with further instructions for returning product that remains in your possession return the concerned products. If you have any questions after reviewing this notice, please call 201-405-1477, ext. 303 between 8:00am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a prompt to leave a voicemail
Quantity in commerce
10 units
Distribution
DuraMatrix-Onlay distributed US Nationwide in the states of FL, IN, Puerto Rico, and the countries of Canada, Chile, and Colombia. RESODURA matrix onlay distributed in Germany.
Product codes and lots
Lot Number 2107292022 Exp. Date: 7/31/2024 UDI: (01)00813954023267(17)240731(10)2107292022 (note: UDI edited 3/7/2022)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0669-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0668-2022Class IITerminatedVoluntary: Firm initiated
Z-0669-2022Class IITerminatedVoluntary: Firm initiated