Z-0659-2022Class I

Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02

Celltrion USA Inc / initiated 2021-12-28 / Terminated / root cause: manufacturing process

CELLTRION USA INC began a recall of Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02 on . FDA classified it as Class I and gives the reason as "The affected lot was placed under quarantine in September 2021 due to complaints of alleged false positives, but 243 kits were erroneously released to customers in December 2021". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0659-2022
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Celltrion USA Inc
Product
Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02
Product code
QKP Coronavirus Antigen Detection Test System.
Reason for recall
The affected lot was placed under quarantine in September 2021 due to complaints of alleged false positives, but 243 kits were erroneously released to customers in December 2021.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initially placed a hold on product due to complaints of false positive and false negative results in September 2021. After product was erroneously distributed from this lot in December 2021, the firm notified affected customers that the lot should not have been distributed and should not be used. This email went out on or around December 28, 2021. Customers were instructed to NOT use the affected lot.
Quantity in commerce
243 kits (6,075 tests)
Distribution
US distribution to CA, FL, NJ, NY, PA, SC.
Product codes and lots
Lot COVGCCM0008

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0659-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0659-2022Class ITerminatedVoluntary: Firm initiated