Z-0659-2022Class I
Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02
Celltrion USA Inc / initiated 2021-12-28 / Terminated / root cause: manufacturing process
CELLTRION USA INC began a recall of Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02 on . FDA classified it as Class I and gives the reason as "The affected lot was placed under quarantine in September 2021 due to complaints of alleged false positives, but 243 kits were erroneously released to customers in December 2021". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0659-2022
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Celltrion USA Inc
- Product
- Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02
- Product code
- QKP Coronavirus Antigen Detection Test System.
- Reason for recall
- The affected lot was placed under quarantine in September 2021 due to complaints of alleged false positives, but 243 kits were erroneously released to customers in December 2021.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initially placed a hold on product due to complaints of false positive and false negative results in September 2021. After product was erroneously distributed from this lot in December 2021, the firm notified affected customers that the lot should not have been distributed and should not be used. This email went out on or around December 28, 2021. Customers were instructed to NOT use the affected lot.
- Quantity in commerce
- 243 kits (6,075 tests)
- Distribution
- US distribution to CA, FL, NJ, NY, PA, SC.
- Product codes and lots
- Lot COVGCCM0008
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0659-2022 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY